Supplemental Characteristics
Measurement
Number of display channels
Four display channels available.
Number of measurement channels
Two primary and two auxiliary measurement channels available.
Measurement parameters
8753ET: Reflection, transmission, A, B, R, A/R, B/R, A/B. Conversion to impedance or admittance.
8753ES: S11, S21, S12, S22, A, B, R, A/R, B/R, A/B. Conversion to impedance or admittance.
Formats
Cartesian: log/linear magnitude, phase, group delay, SWR, real and imaginary.
Smith chart: with log/linear amplitude and phase, R + jX, G + jB, or real/imaginary markers.
Polar: with linear/log amplitude, phase, or real and imaginary markers.
Data markers
Each display channel has five independent markers that can be displayed simultaneously. Twenty inde- pendent markers can be displayed in 4-channel display mode when markers are uncoupled.
Marker functions
Markers can be used in various functions: Marker search (Mkr to max, Mkr to min, Mkr to target), Mkr bandwidth with user-defined target values, Mkr to start, Mkr to stop, Mkr to center, Mkr to span, Mkr to reference, Mkr to delay, and trace sta-
tistics (average value, standard deviation, and peak- to-peak deviation of the data trace between two markers). The tracking function enables continuous update of marker search values on each sweep.
Group Delay Characteristics
Aperture: selectable
Maximum aperture: 20% of frequency span Minimum aperture: (freq. span) / (number of points +1)
Range
The maximum delay is limited to measuring no more than 180° of phase change within the mini- mum aperture. Range = 1/(2 x minimum aperture)
16
Accuracy
The following graph shows group-delay accuracy at 1.3 GHz with type-N full two-port calibration and 10-Hz IF bandwidth. Insertion loss is assumed to be <2 dB and electrical length to be ten meters.
Source Control
Sweep limits
Set start/stop or center/span of the stimulus parameter (frequency, power, time) directly through the source control keys and the control knob, the step keys or the data entry keyboard.
Sweep type
Set a linear or logarithmic sweep, an arbitrarily defined frequency list, a power sweep or a CW (single frequency) type of sweep.
Measured number of points per sweep
Linear frequency: choose 3, 11, 26, 51, 101, 201,
401, 801, or 1601 points.
Fast swept list
Define up to 30 different sub-sweep frequency ranges in any combination of CW, center/span, or start-stop sweep modes. Set test-port power levels and IF bandwidth independently for each segment.
Sweep modes
Set a coupled channel sweep (same stimulus conditions on both channels) or an uncoupled channel sweep.
Chop/alternate
Select whether to alternately or simultaneously (chop) measure channels when in dual-channel mode. Chop mode is faster, while alternate mode optimizes dynamic range.
Sweep time
Set sweep time in seconds, minutes, or hours.
Automatic sweep time
Select auto sweep time by entering zero seconds sweep time. The analyzer will sweep at the minimum sweep time for any subsequently selected stimulus conditions. Auto sweep time is the default condition.
Sweep trigger
Set to either continuous, single, group sweep, or external trigger. Set external trigger to take a com- plete sweep or to measure individual points in a frequency, power or list sweep.
Power
Set source power from -20 to +5 dBm for the
8753ET1 or from -85 to +10 dBm for the 8753ES2. Power slope can be set in dBm/GHz.
Power Meter Calibration
Select continuous leveling or use a correction table to modify source power. The correction table is created with an initial single sweep. Make single or multiple power meter readings at each frequency.
Data Accuracy Enhancement
Measurement calibration
Measurement calibration significantly reduces measurement uncertainty due to errors caused by system directivity, source and load match, tracking, and crosstalk. Full two-port calibration removes all the systematic errors to obtain the most accurate measurements.
Calibration types available
• Frequency response
Simultaneous magnitude and phase correction of frequency response errors for either reflec- tion or transmission measurements.
• Response and isolation
Compensates for frequency response and direc- tivity (reflection) or frequency response and crosstalk errors.
• Enhanced response calibration
Corrects for frequency response and source match for transmission measurements, and provides one-port calibration for reflection measurements.
• One-port calibration
Uses test set port 1 or port 2 to correct for directivity, frequency response and source match errors.
• Two-port calibration
Compensates for directivity, source match, reflection frequency response, load match, transmission frequency response and crosstalk for an S-parameter test set. Crosstalk calibration can be omitted. Available on 8753ES analyzers.
• TRL*/LRM* calibration
Compensates for directivity, reflection and transmission frequency response, and crosstalk in both the forward and reverse directions.
Especially suitable for calibrating non-
coaxial environments, such as in test fixtures. TRL*/LRM* is a special implementation of TRL/LRM calibration, modified for the three- sampler receiver in the 8753ES. Available
on Agilent 8753ES analyzers.
Interpolated error correction
With any type of accuracy enhancement applied, interpolated mode recalculates the error coeffi- cients when the test frequencies are changed. The number of points can be increased or decreased and the start/stop frequencies can be changed, but the resulting frequency span must be a subset of the original calibration frequency span. System performance is not specified for measurements with interpolated error correction applied.
Velocity factor
Enters the velocity factor to calculate equivalent electrical length.
Reference plane extension
Redefine the plane-of-measurement reference to other than port 1 or port 2 of the 8753ET and 8753ES.
Select default calibration kit
Select from a list of standard calibration kits or choose a user-defined kit.
Data Averaging
IF bandwidth
The IF bandwidth is selectable from 6 kHz to 10 Hz.
Weighted sweep-to-sweep averaging
Averages vector data on each successive sweep.
Trace smoothing
Computes the moving average of adjacent data points. Smoothing aperture defines the trace width (number of points) to be averaged, and ranges from 0.25% to 20% of the trace width.
Display Control
Display formats
Single-channel, dual-channel overlay (both traces on one graticule), dual-channel split (each trace on separate graticules), three-channel split (each
trace on separate graticules), three-channel overlay (three traces on one graticule), quad-channel over- lay (four traces on one graticule), quad-channel split (each trace on separate graticules).
1. -85 to +5 dBm with Option 004.
2. +8 dBm maximum with Option 075 or 014. 17
Trace Functions
Display data
Display current measurement data, memory data, or current measurement with measurement and memory data simultaneously.
Trace math
Vector division or subtraction of current linear measurement values and memory data.
Display annotations
Start/stop, center/span, or CW frequency, source level, scale/div, reference level, marker data, softkey functions, warning and caution messages, trace identification, and pass/fail indication.
Autoscale
Automatically selects scale resolution and refer- ence value to center the trace.
Electrical delay
Offset measured phase or group delay by a defined amount of electrical delay, in seconds.
Frequency blanking
Blank out all frequency information on the display. Requires an instrument preset to
re-enable frequency information on the display.
Title
Add custom titles (49 characters maximum) to the display.
Adjust display
Customize the color and brightness of the data traces, memory traces, reference lines, graticules, text, and warning messages. Default colors can
be recalled along with one set of user-defined display values.
Storage
Instrument state
Up to 31 instrument states can be stored internally or recalled via the SAVE/RECALL menu. Instrument states include all control settings, active limit lines, active list frequency tables, memory trace data, active calibration coefficients, and custom display titles. Storage is in nonvolatile memory.
Test sequences
Six measurement sequences can be stored or recalled via the sequencing menu. Sequences may also be recalled from Preset menu. Sequence reg- ister 6 is part of nonvolatile storage and is not erased during a power cycle. If sequence 6 is titled AUTO, it will be executed when power is turned on.
18
Disk drive
Data, instrument states, user graphics, data plots and test sequences can be stored on internal floppy disk in MS-DOS® or Agilent’s standard LIF formats.
Data Hardcopy
Data plotting
Hardcopy plots are automatically produced with HP-GL compatible digital plotters. The 8753ET/ES provides Centronics, RS-232C, and GPIB interfaces.
Data listings
Printouts of instrument data are directly produced with a printer such as the HP DeskJet or LaserJet. Select black & white or color print. For a list of compatible printers, consult our printer compati- bility guide Web page at www.agilent.com/find/pcg
Configure plots
Configure plots completely from the network ana- lyzer by defining pen color and line type for data, text markers, graticules, and memory traces.
Functions
Plot trace(s), graticule(s), marker(s), or text includ- ing operating and system parameters.
Quadrants
Plot entire display in one of four different quad- rants of the plotter paper.
System Capabilities
Limit lines
Define test limit lines that appear on the display
for go/no go testing. Lines may be any combination of horizontal, sloping lines, or discrete data points. Limit-test TTL output available for external control or indication.
External source mode
The receiver (input R) detects and phase-locks to any externally generated CW signal. Receiver inputs A and B will measure this same frequency for comparison or tracking measurements.
Automatic: The input signal frequency is counted and displayed.
Manual: Measures the input signal closest to the fre- quency specified by the user (within +0.5 to +5 MHz).
Tuned receiver
Tunes the receiver for a synthesized CW input signal at a precisely specified frequency. The time bases of the external RF source or sources must be tied to the external reference input (rear panel BNC). The built-in RF source is not used.
Frequency offset on/off
Sets the RF source to be swept at a fixed offset frequency above the receiver as required in a swept RF/IF, fixed LO, mixer test.
Service menu
Select the desired service test, service diagnostic, service or verification mode.
Test Sequences
Description
Create, edit, save, or recall a series of front-panel key- strokes to automate a measurement. Test sequences may contain basic stimulus and measurement func- tions (frequency, power, parameter, format, scale) advanced operations (time domain, limit testing, display marker values) and basic logical branching (IF limit test fails DO sequence 5 or GOSUB).
Storage
Test sequences can be stored internally to a disk drive and can be loaded from a computer over the GPIB interface. Sequence 6 is saved in nonvolatile storage and can be used as an autostart routine when titled AUTO.
General purpose input/output
Read or write bits to the output port to control external devices such as part handlers. Eight output and five input TTL lines are available on the parallel port.
Other functions
PAUSE/continue, wait, title sequence, print sequence, duplicate sequence, pause, and select.
Time-domain (Option 010)
With the time-domain option, data from transmis-
sion or reflection measurements in the frequency domain are converted to the time domain using
a Fourier transformation technique (chirp Z) and presented on the display. The time-domain response shows the measured parameter value versus time. Markers may also be displayed in electrical length (or physical length if the relative propagation veloc- ity is entered).
Time stimulus modes
Two types of time excitation stimulus waveforms can be simulated during the transformations,
a step and an impulse.
• Low-pass step
This stimulus, similar to a traditional time- domain reflectometer (TDR) stimulus wave- form, is used to measure low-pass devices. The frequency-domain data should extend from
DC (extrapolated value) to a higher value.
• Low-pass impulse
This stimulus is also used to measure low-pass devices.
• Bandpass impulse
The bandpass impulse stimulates a pulsed RF signal (with an impulse envelope) and is used to measure the time-domain response of band- limited devices.
Windows
The windowing function can be used to modify (filter) the frequency-domain data and thereby reduce overshoot and ringing in the time-domain response. Three types of windows are available: minimum, normal, and maximum.
Gating
The gating function can be used to selectively remove reflection or transmission time-domain responses. In converting back to the frequency- domain the effects of the responses outside the gate are removed.
Remote Programming
Interface
GPIB interface operates to IEEE 488-1978 and IEC
625 standards and IEEE 728-1982 recommended practices.
Addressing
The GPIB address can be verified or set from the front panel via the local menu and can range from
0 to 30 decimal (factory set at 16).
Pass control
Allows the 8753ET/ES to request control of the GPIB (when an active controller is present) when- ever it needs to output to a plotter or printer.
System controller
Lets an 8753ET/ES become a controller on the
GPIB to directly control a plotter or a printer.
Talker/listener
Lets the 8753ET/ES become a GPIB talker/listener when an external controller is present.
Transfer formats
Binary (internal 48-bit floating-point complex format) ASCII 32- or 64-bit IEEE 754 floating- point format.
User-accessible graphics
Using a subset of HP Graphics Language (HP-GL), vector or text graphics may be written on the
8753ET/ES via GPIB. Up to 5 kbytes of data can be stored at one time (4 bytes per vector, 2 bytes per character).
Interface function codes
SH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP0, DC1, DT0, C1, C2, C3, C10, E2 19
General Characteristics
Front Panel Connectors
8753ES test ports (without Option 011) Connector type 7 mm, precision Impedance 50 ohms (nominal)
8753ES Option 011 test ports
Connector type Type-N female
Impedance 50 ohms (nominal)
8753ES Option 075 test ports
Connector type Type-N female
Impedance 75 ohms (nominal)
8753ET test ports
Connector type Type-N female
Impedance 50 ohms (nominal)
Probe power
+15V ±2% 400 mA
(combined load for both probe connections)
-12.6V ±5.5% 300 mA
(combined load for both probe connections)
Rear-Panel Connectors
External reference frequency input (EXT REF INPUT)
Frequency 1, 2, 5, and 10 MHz
(±200 Hz at 10 MHz)
Level -10 dBm to +20 dBm, typical
Impedance 50 ohms
Connector BNC (f)
High-stability frequency reference output (Option 1D5)
Frequency 10.0000 MHz
Frequency stability ±0.05 ppm
(0° to 55° C)
Daily aging rate <3x10–9/day
(after 30 days)
Yearly aging rate 0.5 ppm/year Output 0 dBm minimum Nominal output impedance 50 ΩConnector BNC (f)
External auxiliary input (AUX INPUT)
Input voltage limits -10V to +10V
External AM input (EXT AM)
±1 volt into a 5 k Ω resistor, 1 kHz maximum,
resulting in approximately 8 dB/volt amplitude modulation.
BNC (f) connector.
External trigger (EXT TRIGGER)
Triggers on a negative TTL transition or contact closure to ground. BNC (f) connector.
20
Test sequence output (TEST SEQ)
By default, this connector outputs a TTL end-of- sweep signal. It can also be programmed by the user in a test sequence to output a user-defined TTL signal. BNC (f) connector.
Limit test output (LIMIT TEST)
This connector outputs a TTL signal of the limit test results. Pass: TTL high. Fail: TTL low. BNC (f) connector.
Test-port bias input (BIAS CONNECT) (8753ES only)
Maximum voltage +30 VDC Maximum current ±200 mA (no degradation in RF specs)
Maximum current ±1 A Connector BNC (f)
VGA video output (EXT MON)
This connector drives external VGA monitors.
GPIB
This connector allows communications with compatible devices including external controllers, printers, plotters, disk drives, and power meters.
Parallel port
This 25-pin female connector is used with parallel peripherals. It can also be used as a general purpose I/O port.
RS-232C
This 9-pin male connector is used with serial peripherals.
DIN keyboard
This mini-DIN connector is used for adding an
IBM PC-AT compatible keyboard.
Test set interconnect
This connector is used to connect an 8753ES Option 011 to the 85046A/B or 85047A test set.
8753ES analyzers without Option 011 can use signal levels on this connector for sequencing or general purpose I/O applications.
Internal memory
Typical data retention time with 3V, 1.2 Ah battery:
At 25° C 11,904 days (32.6 years) At 40° C 1244 days (3.4 years)
At 70° C 250 days (0.68 year)
Line power
48 Hz to 66 Hz
115V nominal (90V to 132V) or 230V nominal (198V to 264V).
280 VA max.