The Stanford Research Systems (SRS) DS345 Function and Arbitrary Waveform Generator is a full-featured, 30 MHz instrument that uses an innovative Direct Digital Synthesis (DDS) architecture. It generates many standard waveforms with excellent frequency resolution (1 µHz), including: sine waves and square waves at frequencies up to 30.2 MHz, and triangle and ramp waveforms up to 100 kHz. In addition to the standard waveforms, the unit also provides a wideband (10 MHz) white noise source and has versatile modulation capabilities including AM, FM, Burst, PM and frequency sweeps. It also generates arbitrary waveforms with a fast 40 Msample/s update rate and 12-bit vertical resolution. A 10 MHz rear-panel input allows the DS345 to be synchronized to an external timebase. A 10 MHz rear-panel output allows multiple DS345's to be phase locked.

Features and Specifications of the SRS DS345 Function and Arbitrary Waveform Generator include:

  • 1 µHz to 30.2 MHz frequency range

  • 1 µHz frequency resolution

  • Sine, square, ramp, triangle and noise

  • Phase continuous frequency sweeps

  • AM, FM, burst and phase modulation

  • 16,300 point arbitrary waveforms

  • 10 MHz reference input

  • Optional RS-232 and GPIB interfaces

  • Frequency Range:

    • Sine:

      • Max Frequency:  30.2 MHz

      • Resolution:  1 µHz

    • Square:

      • Max Frequency:  30.2 MHz

      • Resolution:  1 µHz

    • Ramp:

      • Max Frequency:  100 kHz

      • Resolution:  1 µHz

    • Triangle:

      • Max Frequency:  100 kHz

      • Resolution:  1 µHz

    • Noise:

      • Max Frequency:  10 MHz

      • Resolution:  (Gaussian weighting)

    • Arbitrary:

      • Max Frequency:  10 MHz

      • Resolution:  40 MHz/n (sample rate)

  • Output:

    • Source Impedance:  50 O

    • Grounding:  Output may float up to ±40 V (AC + DC) relative to earth ground

  • Amplitude:

    • Range:               0.01 to 10 Vpp (50 O), 20 Vpp (Hi-Z)

    • Resolution:          3 digits (DC offset = 0 V)

    • Sine Wave Accuracy (0 V DC offset):

      • 5 to 10 Vpp:

        • ±0.2 dB (1 µHz to 20 MHz)

        • ±0.3 dB (20 MHz to 30.2 MHz)

      • 0.01 to 5 Vpp:

        • ±0.4 dB(1 µHz to 20 MHz)

        • ±0.5 dB (20 MHz to 30.2 MHz)

    • Square Wave Accuracy:

      • 5 to 10 Vpp:

        • ±3 % (1 µHz to 100 kHz)

        • ±6 % (100 kHz to 20 MHz)

        • ±15 % (20 MHz to 30.2 MHz)

      • 0.01 to 5 Vpp:

        • ±5 % (1 µHz to 100 kHz)

        • ±8 % (100 kHz to 20 MHz)

        • ±18 % (20 MHz to 30.2 MHz)

    • Triangle, Ramp, Arbitrary Accuracy:

      • ±3 % (>5 Vpp)

      • ±5 % (<5 Vpp)

  • DC Offset:

    • Range:  ±5 V (limited such that |VAC peak| + |VDC| < 5 V)

    • Resolution:  3 digits (VAC = 0)

    • Accuracy:

      • 1.5 % of setting + 0.2 mV (DC only)

      • ±0.8 mV to ±80 mV, depending on AC and DC settings

  • Sine Wave:

    • Spurious Components:  <-55 dBc (non-harmonic)

    • Phase Noise:  <-50 dBc in a 30 kHz band centered on the carrier, exclusive of discrete spurious signals

    • Sub-Harmonic:  <-50 dBc

    • Harmonic Distortion:

      • Level: <-55 dBc / Freq Range: DC to 100 kHz

      • Level: <-45 dBc / Freq Range: 0.1 to 1 MHz

      • Level: <-35 dBc / Freq Range: 1 to 10 MHz

      • Level: <-25 dBc / Freq Range: 10 to 30 MHz

  • Square Wave:

    • Rise/Fall Time:  <15 ns (10 to 90 %), at full output

    • Asymmetry:   <1 % of period + 4 ns

    • Overshoot:   <5 % of peak to peak amplitude at full output

  • Ramps, Triangle and Arbitrary Waveforms:

    • Rise/Fall Time:  45 ns (10 MHz Bessel filter)

    • Linearity:  ±0.5 % of full-scale output

    • Settling Time:  <1 µs to settle within 0.1 % of final value at full output

  • Arbitrary Waveforms:

    • Sample Rate:  40 MHz/N, N = 1 to 234-1

    • Memory Length:   8 to 16,300 points

    • Resolution:   12 bits (0.025 % of full scale)

  • Phase:

    • Range:  ±7199.999° with respect to arbitrary starting phase

    • Resolution:  0.001°

  • Amplitude Modulation:

    • Source:  Internal (sine, square, triangle or ramp) or External

    • Depth:  0 to 100 % AM or DSBSC

    • Rate:  0.001 Hz to 10 kHz internal, 15 kHz max external

    • Distortion:  <-35 dB at 1 kHz, 80 % depth

    • DSB Carrier:  <-35 dB typ at 1 kHz modulation rate (DSBSC)

    • External Input:  ±5 V for 100 % modulation, 100 kO impedance, 15 kHz bandwidth

  • Frequency Modulation:

    • Source:  Internal (sine, square, triangle, ramp or arbitrary)

    • Rate:  0.001 Hz to 10 kHz

    • Span:  1 mHz to 30.2 MHz (100 kHz for triangle, ramp)

  • Phase Modulation:

    • Source:  Internal (sine, square, triangle, ramp)

    • Rate:  0.001 Hz to 10 kHz

    • Span:  ±7199.999°

  • Frequency Sweep:

    • Type:  Linear or log, phase continuous

    • Waveform:  Up, down, up-down, single sweep

    • Time:  0.001 s to 1000 s

    • Span:  1 mHz to 30.2 MHz (to 100 kHz for triangle, ramp)

    • Markers:  Two markers may be set at any sweep point (TTL output)

    • Sweep Output:  0 to 10 V linear ramp signal, synchronized to sweep

  • Burst Modulation:

    • Waveform:  Any waveform except noise may be burst modulated

    • Frequency:  Sine, square to 1 MHz
      Triangle, ramp to 100 kHz
      Arbitrary to 40 MHz sample rate

    • Count:  1 to 30,000 cycles/burst (1 µs to 500 s burst time limits)

  • Trigger Generator:

    • Source:  Single, Internal, External, Line

    • Rate (Internal):  0.001 Hz to 10 kHz (2-digit resolution)

    • External Trigger:  Positive or negative edge, TTL

    • Output:  TTL level

  • Standard Timebase:

    • Accuracy:  ±5 ppm (20 °C to 30 °C)

    • Aging: 5 ppm/year

    • Input:  10 MHz/N ± 2 ppm (N = 1 to 8) 1 Vpp minimum input level

    • Output:  10 MHz, >1 Vpp sine into 50 O

  • General:

    • Interfaces:  Optional RS-232 (300 to 19.2 kbaud, DCE) and GPIB with DOS based arbitrary waveform software (AWC). All instrument functions are controllable over the interfaces.

    • Non-Volatile Memory:  Nine sets of instrument settings can be saved and recalled

    • Power:  50 W, 100/120/220/240 VAC, 50/60 Hz